Regulatory Open Forum

 View Only
  • 1.  Sterility testing

    Posted 28-Apr-2016 11:21
    Hi,
    Is sterility testing a requirement for autoinjectors?
     
    Thanks!
    Betzy


  • 2.  RE: Sterility testing

    Posted 28-Apr-2016 11:54
    Yes, Betzy!

    For both autoinjector itself and the finished product (device/biologics or device/drugs).

    You may refer to FDA technical guidance in 2013

    D





  • 3.  RE: Sterility testing

    Posted 29-Apr-2016 07:42

    Betzy,

     

    Reusable autoinjectors are NOT required to be sterile or sterilized.  Disposable autoinjectors, which are provided with the drug (usually a Prefilled syringe)  are also not required to be sterile except that the syringe or self-contained primary container closure must maintain the sterility of the injectable solution.  The only exception is where the design of the disposable autoinjector is such that the injector contains part of the fluid path that contacts the drug.  These designs usually use cartridges instead of Prefilled syringes that are punctured by a needle in the injector, which must of course be sterile..

     

    Lee Leichter

    President

    P/L Biomedical

    10882 Stonington Avenue

    Fort Myers, FL 33913 USA

    O: +1-239-244-1448

    F: +1-815-550-0162

    C: +1-239-994-6488

    E: leichter@plbiomedical.com