Michael,
It seems you are in Brazil. I suppose you are following Br. Official Gazette, dated 2/1/2010 for the type of infusion pump you are working on. There are some differences as for the use of the standards (e.g., gravity drip infusion equipment/pump) between UK, Brazil, and US.
In view of ISO 8536-8 and IEC 60601-2-24, my understanding has been that:
As for testing temp., it should be done at 20C as the intended use reflects.
As for the infusion sets per ISO 8536-8, 40C should be in your tech spec.
Thanks!
-------------------------------------------
President and Principal
http://www.regulatorydoctor.com Riner VA
United States
-------------------------------------------
Original Message:
Sent: 09-12-2013 08:20
From: Michael Dun
Subject: Conflicting Infusion Set/Pump Requirements
Hi Jan,
Thank you for your response on my posting. Your feedback is logical and aligns with the risk-based strategy that I had planned on adopting in this scenario. The intended use and design of the pump and infusion sets should assist me in determining the impact that temperature will have upon storage precision limits and therefore which temperature to use during testing. As to the specific difference in the standards I've also reached out to ISO for clarification and should hopefully receive a response from them soon.
Best regards,
Michael
-------------------------------------------
Michael Dun
Campinas
Brazil
-------------------------------------------
Original Message:
Sent: 09-11-2013 11:05
From: Jan Peterson
Subject: Conflicting Infusion Set/Pump Requirements
While I can see the technical problem of resolving these two discordant standards, the actual difference with regard to a reservoir capacity should be trivial (<0.5% volume difference of water between 20° and 40° C). This is a reservoir and you want to know and label its capacity properly; this is not a volumetric flask trying to meet some high standard of precision. If the reservoir was expected to hold 1000 ±10 mL of water, it would not make any significant difference if this were measured at 20° or 40°, as you would still be well within the ±10 mL range. It would be even less different if the reservoir were for saline, as that coefficient of expansion is smaller. My point being, either method will get you what you need, just do not set an unnecessarily tight design tolerance for the capacity of the resevoir. Better to attend to the delivery rates of the pumping system, where precision is more important. If your pump was designed to warm the reservoir closer to body temperature prior to infusion, then it might make sense to use the higher temperature as the reference. But why the standards seem so different, I cannot say.
-------------------------------------------
Jan Peterson RAC
Rockville MD
United States
-------------------------------------------
Original Message:
Sent: 09-10-2013 09:43
From: Michael Dun
Subject: Conflicting Infusion Set/Pump Requirements
Dear RAPS Community,
Is anyone able to help provide some clarification/assistance regarding what appear to be conflicting requirements for infusion set/pump performance and labeling requirements as described in ISO 8536-8 and IEC 60601-2-24?
Specifically, ISO 8536-8, under 6.14 describes that storage volume is to be determined in accordance with IEC 60601-2-24 and declared in labeling in accordance with ISO 8536-8 10.1(h). However IEC 60601-2-24 describes that the testing is to be performed at 20C +/- 2C / 65%RH +/- 5%RH, while ISO 8536-8 10.1(h) describes that the storage volume should be stated at 40C.
Any assistance on this would be greatly appreciated.
Best regards,
Michael
-------------------------------------------
Michael Dun
RA & Technical Consultant
Ambiente & Qualidade
Campinas
Brazil
-------------------------------------------