820.70(a) applies to inspection, measuring, and test equipment (IM&TE). When you say they did validation, I infer you mean process validation under 820.75. In this case, one controls the process input parameters to provide a high degree of assurance that the process output, the product, is conforming. The reason for this approach is that the product cannot (or is not) fully verified by subsequent inspection or test.
Process validation set the acceptable values of the input parameters: time, temperature, and pressure in your example. Implicit is that one knows the actual value of these parameters during process validation, at setup for a production run, and during the run. 820.75(b) requires monitoring and recording of the data.
Consequently, these parameters must be measured, and so IM&TE is involved. Often, in injection molding, the IM&TE is built into the molding machine. For example there may be a pressure sensor with a display on the operating screen for the machine. The pressure sensor and display are IM&TE and subject to 820.70(a). The section equally applies to IM&TE built into production equipment and to stand-alone IM&TE. It is the existence of the measuring function that invokes 820.75(a).
FDA would not accept process validation or product verification in lieu of calibration, since they address very different issues. In the conventional process validation model, the IQ step would ensure that any IM&TE is in the calibration program. In addition, the OQ and PQ portions typically require that the IM&TE is within its calibration interval at the start and end. If not, you won't get valid results. In addition, the IM&TE must be in the calibration interval for any production runs.
Based on your description, the 483 citation is valid. You response should be a project to identify all IM&TE in the plant; establish the procedures to ensure that equipment is routinely calibrated, inspected, checked, and maintained; and bring all of the IM&TE under the aegis of the procedure.
In my opinion, you should also expect a Warning Letter.
-------------------------------------------
Dan O'Leary
Swanzey NH
United States
-------------------------------------------
Original Message:
Sent: 06-25-2015 07:40
From: Bethany Hills
Subject: Calibration Procedures - Needed?
I am working with a client that uses an injection mold process and a sonic weld in manufacturing medical device. They have an FDA 483 inspection report that is requiring them to calibration the instruments. Since they haven't done calibration before (only validation and verification), they are worried that calibration would impact their validation. They are looking for some help understanding what others are doing - verifying time and temp and pressure? Would FDA accept verification plan in place of calibration?
I told them I'd reach out to my contacts that would have injection molding and get some thoughts. The FDA is citing 820.72(a) and saying they need to establish procedures to ensure equipment is routinely calibrated, inspected, checked and maintained...Does this mean they must calibrate a piece of equipment or only aspects that need calibration (like gauges, etc).
-------------------------------------------
Bethany Hills JD
Attorney
Epstein Becker & Green PC
New York NY
United States
-------------------------------------------