It is not clear what you mean by calibration records, so I will assume these are calibration certificates and data packages from an external supplier of calibration services. You then ask about the need to keep electronic records. In this I would infer you propose keying in the results from the supplier.
If you scan them in, then you are creating a copy. I would be no different than keeping a copy on paper using a copy machine. Many years ago, you might have captured the image to microfilm. More recently, but still in the “dark ages”, you might have use 8” floppy disks. In all these cases the same principle applies, you have captured an image on some medium.
The next problem is storage. You need to define the record retention time (the storage time) as well as criteria to preserve the record. On paper you need to consider moisture, acid content of the paper, etc. For the floppy disks you need to consider magnetic fields, physical configuration (so they don’t warp over time), etc.
The next problem is retrieval. You should be able to read the record at any time during the record retention time. If you used 8” floppy disks you might have a problem today unless you migrated the image to newer media.
Mehul gave you some record retention requirements, but they were not complete. In FDA QSR the requirement is “for a period of time equivalent to the design and expected life of the device, but in no case less than 2 years from the date of release for commercial distribution by the manufacturer”. In the medical device directive, Annex II the requirement is “for a period ending at least five years, and in the case of implantable devices at least 15 years, after the last product has been manufactured”.
I find it useful for a record retention procedure to include both when the retention period starts and the length of the period. Then, the actual records should be marked with, at least, both items to help prevent premature disposal. I have also seen companies specify the disposal method. For example, sensitive information stored in paper records should be shredded, not put in a dumpster.
------------------------------
Dan O'Leary
Ombu Enterprises, LLC
Swanzey NH
United States
Original Message:
Sent: 12-08-2015 09:28
From: Keri Froese
Subject: Electronic Records
We have started migrating to electronic records for the things that we can, which I think is great. I have a question about calibration records, if I scan these do I need to keep an electronic copy from a regulatory/quality point of view?
Keri