Anon,
Lexamed, Steris and NAMSA also do this work, as do many others besides Nelson labs. Richard is correct that it is best to get help if you are not experienced in the details.
You didn't identify the type of your device, but labs usually ask for the whole finished device or representative prototype to look it over, and to evaluate its materials and surface finishes, to develop an initial protocol.. Sometimes labs suggest total immersion, which is crazy for an electronic device where only its surface contacts the user or patient. So if you give them a complete device, you might make it only a shell, if electronics are sealed on the inside and in use a customer wouldn't be immersing it. Discuss this with a lab.
I would recommend getting at least 3 competitive quotes, because prices can vary incredibly. But be sure you are asking for the same scope of work and exact same protocol when you bid it out, and that the lab has experience. Do not just let them run independently. Assign someone in your organization to ride herd on the lab, and ask for draft protocol and draft reports for your review before they are finalized.....you will be much happier if you don't just throw it "over the wall".
Good luck!
------------------------------
Ginger Cantor, MBA, RAC
Founder/Principal Consultant
Centaur Consulting LLC
River Falls, Wisconsin 54022 USA
715-307-1850
centaurconsultingllc@gmail.com------------------------------
Original Message:
Sent: 25-Apr-2020 02:45
From: Richard Vincins
Subject: cleaning/disinfection validation for non-critical devices
Hello Anon,
Unless you are quite familiar with cleaning and disinfecting validations yourself, it is always wise to get expert assistance on this. There are a number of labs conducting this type of testing like Nelson Laboratories. These labs assist in the review of your device to properly set out a cleaning validation which you are correct using soils to put on the device and then testing the cleaning process. This is a fairly well established testing protocol as done for many other type of devices to show they are "clean." The disinfecting protocol depends on the level of disinfection you are claiming and the type of disinfecting that are used or more precisely can be used. There are many different types of disinfectants and depending on the needs of your device could be just wipes or solutions to use; also need to consider any damage may occur from disinfectants because some materials have adverse reactions to the chemicals. This can be quite an involved activity depending on your device, so it would be good to obtain expert support either in guidance or actual testing.
------------------------------
Richard Vincins RAC
Vice President Global Regulatory Affairs
Original Message:
Sent: 24-Apr-2020 11:36
From: Anonymous Member
Subject: cleaning/disinfection validation for non-critical devices
This message was posted by a user wishing to remain anonymous
Hi all,
we are a small company and have a non-critical device (intact skin contact) that can be reused by multiple patients. So far we have been addressing cleaning and disinfection in a very basic way (i.e., by using disinfection products commonly used in healthcare environments).
Reading the FDA guidance on reprocessing, it looks like we need to conduct validation activities to demonstrate: 1) that the methods (manual in our case) are adequate to allow the device to undergo further processing and to eventually be reused safely; and 2) that the reprocessing instructions are effective in conveying the proper reprocessing methods to the user.
I am looking for advice on how to develop cleaning and disinfection validation protocols that are not too burdensome and how to perform testing. Do you recommend to involve testing labs to examine soil residuals and bacteria presence even for non-critical devices?
Thank you