Konrad-
I'm not sure if you're referring to device or drugs. I'll answer for devices.
You would have to look at the ISO, ASTM, IEC etc standards that you are using to identify such acceptance criteria. There are howvere FDA guidance documents for specific classes of devices (drugs also I would assume) that list safety and performance requirements which would directly point to what standards you should be using. Here's the link :
https://www.fda.gov/RegulatoryInformation/Guidances/Additionally, you can look up previous submissions (510(k)s if you are class 2 device manufacturer) on devices similar to yours from the FDA 510(k) database and identify what standards the industry generally complies with, this could save you time but is dependent on degree of novelty of your device which may encourage additional standardized testing.
Hope this helped.
------------------------------
Vidyalakshmi Jayaraman
Regulatory Affairs Specialist
Marlborough MA
United States
------------------------------
Original Message:
Sent: 05-Dec-2017 11:38
From: Konrad Kulak
Subject: assay validation - when a parameter is validated?
Dear All,
Is there FDA guide that would clearly specify percentage of samples that have to pass the acceptance criteria in order for a parameter (in assay validation) to be considered 'validated' (for example, 75% of samples have to pass the acceptance criteria so a parameter is validated)?
Thank you,
Konrad
------------------------------
Konrad Kulak
Biostat Solutions, Inc.
Frederick MD
United States
------------------------------