Hi,
The FDA guidance on the 'Use of ISO 10993-1' is really helpful, but in your case I would consider first reading ISO 10993-1, especially version 2018.
The scope of the standard (section 1) tells you: This document applies to MDs that have direct or indirect contact with
- the patient's body during intended use;
- the user's body, if the MD is intended for protection (...)
If your risk assessment identifies possible biological adverse effects for the user, of course you could use ISO 10993-1 as well for risk mitigation.
But it may be worth to double check if ISO 10993-1 really applies to your device. There is also an additional note in section 5.2.2 how to deal with devices that come into contact with the users ungloved hands.
As far as I know E&L testing can not be used to address local effects like sensitization or irritation. I would love to learn something different from the RAPS crowd because I personally do not like in-vivo tests that much.
Based on the biological risks you identify in combination with the table in Appendix A you can define the endpoints you need to consider further on. From there you can dive into other standards from the 10993-series.
Read ISO 10993-1 carefully: It tells you in various sections when to consider specific biological endpoints, e.g. section 6.1 tells you when to consider degradation products.
Kind regards,
Michael
------------------------------
Michael Hottner
Köln
Germany
------------------------------
Original Message:
Sent: 19-Feb-2019 04:57
From: Anonymous Member
Subject: extractables and leachables
This message was posted by a user wishing to remain anonymous
Hi all,
When is test on extractables and leachables needed?
In the case of an external device that has surface skin contact (<24 h) only with the practitioner (not the patient), I know from the FDA guidance that cytotoxicity, Irritation and skin sensitization apply.
I don't have clear thought if the following apply as well:
EN/ISO 10993- 13 Biological evaluation of medical devices - Part 13: Identification and quantification of degradation products from polymeric medical devices
EN/ISO 10993- 16 Biological evaluation of medical devices - Part 16: Toxicokinetic study design for degradation products and leachables
EN/ISO 10993- 17 Biological evaluation of medical devices - Part 17: Establishment of allowable limits for leachable substances
EN/ISO 10993- 18 Biological evaluation of medical devices – Part 18: Chemical characterization of materials
Please advise,
Thanks