Regulatory Open Forum

 View Only
  • 1.  Is Usability Testing required for IVDs?

    This message was posted by a user wishing to remain anonymous
    Posted 23-Sep-2022 09:04
    This message was posted by a user wishing to remain anonymous

    Is there anything that would make usability testing required for regulatory compliance for an IVD intended for professional use? Or is it a best practice? 

    Thank you in advance.


  • 2.  RE: Is Usability Testing required for IVDs?

    Posted 23-Sep-2022 11:30
    Edited by Kevin Randall 23-Sep-2022 11:39
    In short, if the nature and complexity of the device, in combination with the users' attributes somehow affects the usability of the device, then appropriate usability design and testing are needed.  Remember that a key goal of usability design and testing is generally aimed at reducing human error.  Often, such use error isn't a function of training or education, but rather is caused by biological human attributes, like size, strength, inability to see in the dark, etc.

    From a more standardized/formal perspective, think about things like the fact that EN/IEC 62366-1 (as amended) for example requires the manufacturer to establish a usability engineering process to provide safety for the patient, user, and others.  It defines "user" as any person interacting with (i.e., operating or handling) the device, and says that common users include clinicians.  In other words, it requires proper consideration and design to accommodate the user's usability needs, even when the user is a trained clinician.

    AAMI HE75 is another example of a standardized paradigm that demands design to meet the user's usability needs, even if the user is a trained clinician.

    ------------------------------
    Kevin Randall, ASQ CQA, RAC (U.S., Europe, Canada)
    Principal Consultant
    Ridgway, CO
    United States
    © Copyright 2022 by ComplianceAcuity, Inc. All rights reserved.
    ------------------------------



  • 3.  RE: Is Usability Testing required for IVDs?

    Posted 23-Sep-2022 11:38
    By the way, AAMI HE75 also indicates that usability design and testing are expected for IVD devices and non-IVDs alike.

    Another interesting example is that Europe's IVDR expects design for usability of IVDs.  For example, see IVDR GSPR 5.

    We could undoubtedly find many other organic examples showing that design for usability is expected for IVDs and non-IVDs alike.

    ------------------------------
    Kevin Randall, ASQ CQA, RAC (U.S., Europe, Canada)
    Principal Consultant
    Ridgway, CO
    United States
    © Copyright 2022 by ComplianceAcuity, Inc. All rights reserved.
    ------------------------------