    Hi,
    I know that FDA defined an expiry date format of YYYY-MM-DD in the UDI Regulation (21 CFR 801.18).  As an aside, I wish they would allow alternates of YYYY-MM and YY-MM, or at least apply enforcement discretion. 

    However my question concerns the MDR.  After reviewing the MDR and the MDCG guidances on UDI, I can find no explicit description of the format for the expiration date with the exception of Annex I (GSPR) where it is stated an expiry date should be stated at least in terms of year and month [23.3(i)].  To add confusion, for date of manufacture [23.3(h)] the term month and year is used.
    Is anyone aware of any definitive format required (or recommended) for EU expiry dates going forward?
    Thanks,
    James


    ------------------------------
    James Bonds J.D.
    Director Regulatory Affairs
    Atlanta GA
    United States
    ------------------------------


  • 2.  RE: Expiry Date Format for MDR

    Posted 25-Jun-2020 10:10

    In terms of the FDA, the date format on labels is not part of UDI. However, the change in the regulation is in the UDI rule. The problem was that the date format without the day was ambiguous. For example, an expiration date "June 2020" could mean June 1, 2020 or June 30, 2020 or some other day in June.

    Initially FDA proposed a different date format, but received many comments that it would not satisfy the ISO standard.

    In one place the preamble says, "FDA received many comments (approximately 110) on the proposed date format and the proposed compliance date of the date format. Nearly all of these comments opposed the proposed date format, considered the time provided to implement this labeling change to be inadequate, or both. Although many comments recognized the benefits of standardized dates, most viewed FDA's proposal as too restrictive, too burdensome, inconsistent with the needs of international commerce, and inconsistent with existing industry practices. Comments noted that FDA's proposed date format would require different labels for a device when marketed in the United States and the same device when marketed in the European Union or other international markets. For example, comments noted that the date format required by the proposed rule "is not consistent with global requirements" and "perpetuates an opportunity for confusion" by not implementing "standardized international dating systems." "

    In another place the preamble says, "One comment suggested that FDA should permit a manufacturer to use any date format it chooses, "as long as the manufacturer makes clear" what format it is using, and a similar comment suggested FDA should "should allow for multiple data formats" but should give "priority…to international standards." Several comments suggested that FDA should permit truncated dates, using only the year and month (YYYY-MM). This is one of the formats permitted under some international standards, such as International Organization for Standardization (ISO) 8601:2004, that were cited by comments." FDA disagreed with this comment.

    In terms of the EU, this is an area in flux. There is EN 1041:2008 Information supplied by the manufacturer of medical devices. If I recall correctly it includes references to the date formats. However, ISO 20417 Medical devices – Information to be supplied by the manufacturer is planned for publication in June 2020. I anticipate that CEN will adopt it shortly thereafter.

    In any case, I recommend using it to show compliance with the MDR Annex I.



    ------------------------------
    Dan O'Leary CQA, CQE
    Swanzey NH
    United States
    ------------------------------



  • 3.  RE: Expiry Date Format for MDR

    Posted 26-Jun-2020 07:08
    Hello James

    Yes, unfortunately EU has an indirect requirement in EN/ ISO 15223-1 where the manufacturing symbol specifics  a date  format YYYY-MM , and this symbol is called out in section 7.2.2 I believe of IEC 60601-1...

    (Sigh).. they conflict.   Personally I think not having flexibility in exact  date format is   if you can explain it elsewhere. 


    ------------------------------
    Ginger Cantor, MBA, RAC
    Founder/Principal Consultant
    Centaur Consulting LLC
    River Falls, Wisconsin 54022 USA
    715-307-1850
    centaurconsultingllc@gmail.com
    ------------------------------



  • 4.  RE: Expiry Date Format for MDR

    Posted 26-Jun-2020 15:37

    I looked at ISO 15223-1:2016.

    For the date of manufacture section 5.1.3 says, "This symbol shall be accompanied by a date to indicate the date of manufacture. This shall be expressed as in ISO 8601 as four digits for the year and, where appropriate, two digits for the month and two digits for the day."

    For the expiration date section 5.1.4 says, "This symbol shall be accompanied by a date to indicate that the medical device should not be used after the end of the year, month or day shown. The date shall be expressed as in ISO 8601 as four digits for the year and, where appropriate, two digits for the month and two digits for the day."



    ------------------------------
    Dan O'Leary CQA, CQE
    Swanzey NH
    United States
    ------------------------------



  • 5.  RE: Expiry Date Format for MDR

    Posted 26-Jun-2020 15:56
    Thanks Dan

    Maybe it's the where appropriate issue.

    Have a great weekend!

    ------------------------------
    Ginger Cantor, MBA, RAC
    Founder/Principal Consultant
    Centaur Consulting LLC
    River Falls, Wisconsin 54022 USA
    715-307-1850
    centaurconsultingllc@gmail.com
    ------------------------------



  • 6.  RE: Expiry Date Format for MDR

    Posted 27-Jun-2020 03:35
    The DIS 15223-1:2020 will likely drop the text that states in the Additional requirements column in regards to the date details (i.e. "In Europe:

    - the date could be a year, year and month, or year, month and day, as required in the relevant EU Directive;"). The primary reason for this was to delete references to national requirements.  National requirements belong in national deviations of in the European version in the Annex Zs of the standard. The HAS consultant has been involved in the process and there is a team working on the Annex Zs.  The DIS 15223-1:2020 (that is the team's hope) is likely to go to ISO to get finalized in the next several weeks or so.  From there we are estimating about 2 - 4 months til publication.  Will advise when know more.

    PS - I have recently joined the ISO TC 210 WG 3 team and got to work thru the 400+ comments received.



    ------------------------------
    Leonard (Leo) Eisner, P.E.
    The "IEC 60601 Guy"
    Principal Consultant, Eisner Safety Consultants
    Phone: (503) 244-6151
    Mobile: (503) 709-8328
    Email: Leo@EisnerSafety.com
    Website: www.EisnerSafety.com
    ------------------------------