  • 1.  Biocompatibility Coupon

    Posted 06-Feb-2018 15:17

    Does anyone know how to estimate the size of a coupon for biocompatibility testing representative of an implantable medical device?


    Globus Medical Inc Lori Burns  | Director, Regulatory Affairs

    Globus Medical, Inc.
    Valley Forge Business Center
    2560 General Armistead Avenue | Audubon, PA 19403
    Ph: (610) 930-1800 ext. 2950 | Fax: 610-930-2042
    An ISO 13485 Registered Company
     
    website | map | email   
     
    Confidentiality Note:  This email is confidential and intended solely for the use of the individual to whom it is addressed. If you are not the intended recipient, be advised that you have received this email in error and that any use, dissemination, forwarding, printing, or copying of this email is strictly prohibited. If you have received this email in error please contact the sender. Any views or opinions presented are solely those of the author and do not necessarily represent those of Globus Medical, Inc. Although this email and any attachments are believed to be free of any virus or other defects which might affect any computer or IT system into which they are received, no responsibility is accepted by Globus Medical, Inc. for any loss or damage arising in any way from the receipt or use thereof.


  • 2.  RE: Biocompatibility Coupon

    Posted 06-Feb-2018 15:32
    ​Have you looked at Annexes B, C, and D of ISO 10993-6?

    ------------------------------
    Kevin Randall, ASQ CQA, RAC (U.S., Canada, Europe)
    Principal Consultant
    ComplianceAcuity, Inc.
    Golden CO
    United States
    www.complianceacuity.com
    ------------------------------



  • 3.  RE: Biocompatibility Coupon

    Posted 07-Feb-2018 12:17
    Work with a testing lab to determine the best specs. There isn't a predetermined size, it depends on the contact, the test, and the levels of exposure desired. But, without question, the coupon has to have all of the material, in roughly proportionate volumes and all final processing to be considered valid for biocompatibility testing.

    ------------------------------
    Lisa Olson
    RCRI - ENTERPRISE
    Minneapolis MN
    United States
    ------------------------------