  • 1.  sampling plan for verification activities

    This message was posted by a user wishing to remain anonymous
    Posted 30-Sep-2020 11:41
    This message was posted by a user wishing to remain anonymous

    Hello,
    Can somebody suggest a good reference (e.g., possible metholodologies) for choosing the number of units that shoud be used for verification testing? Example of testing is:
    - electrical safety or emc according to 60601-1 and 60601-1-2
    - fatigue testing
    Right now the best justification I have is that "the device is extremely expensive", which I don't think will work very well with the FDA reviewer.
    Please let me know if you have any suggestions on guidances, standards, books, etc. that can help us build justifications for testing on a small numbers of units (or even one unit, as in the case of 60601 series).
    Thanks


  • 2.  RE: sampling plan for verification activities

    Posted 30-Sep-2020 13:24
      |   view attached
    Hello,

    I am not an expert in 60601-testing, but I have seen mainly 'type-testing' (1 device) in combination with final production acceptance tests so far.
    Generally statistics play an important role in verification testing. The higher the associated risk level, the higher the amount of data points and/or samples you will need.
    Check out the paper attached to my reply.

    Best wishes,
    Michael

    ------------------------------
    Michael Hottner
    Köln
    Germany
    ------------------------------



  • 3.  RE: sampling plan for verification activities

    Posted 01-Oct-2020 10:29
    Are you able to characterize the expected/potential variability in the manufacturing process?​

    ------------------------------
    Anne LeBlanc
    Manager, Regulatory Affairs
    United States
    ------------------------------



  • 4.  RE: sampling plan for verification activities

    This message was posted by a user wishing to remain anonymous
    Posted 01-Oct-2020 12:18
    This message was posted by a user wishing to remain anonymous

    Hi Anne, 
    unfortunately not, we are behind in this type of activity. 
    Any suggestion you can provide on how to assess the variability would be very helpful.

    Michael, thank you a lot for the reference!



  • 5.  RE: sampling plan for verification activities

    Posted 09-Oct-2020 09:29
    We use ANSI/ASQ Z1.4/Z1.9. I've never had an auditor raise an issue to those.

    ------------------------------
    Jeffrey Freedman
    QA/RA Specialist III
    Lowell MA
    United States
    ------------------------------