  • 1.  UDI for single-use implantable device

    Posted 20-Mar-2022 02:16
    Hello all,

    I would like to know if there is any requirement to bear a UDI barcode on an single-use implantable device (on the device itself) and not only on the package. 

    Any comment would be appreciated,
    Thanks
    Assaf Yichia


    ------------------------------
    Assaf Yichia
    QA Manager
    Yokneam Illit
    Israel
    ------------------------------


  • 2.  RE: UDI for single-use implantable device

    Posted 20-Mar-2022 10:18
    Hello Assaf

    I'm not aware of any such requirement in Israel. Generally, countries that follow the IMDRF model require the UDI on the packaging where it can be read prior to the implantation, and no direct marking. (IMDRF/UDI WG/N48 FINAL: 2019)

    ------------------------------
    Anne LeBlanc
    United States
    ------------------------------



  • 3.  RE: UDI for single-use implantable device

    Posted 20-Mar-2022 10:30
    Thank you Anne for your answer.

    pPer section 12.2 (IMDRF/UDI WG/N48 FINAL: 2019) - The IMDRF UDI Guidance (IMDRF/WG/N7Final:2013) reusable should have a UDI Carrier on the device itself
    Is this still a requirement for reusable device? Is there any requirement for single-use device also? our market will be USA.

    thanks
    Assaf


    ------------------------------
    Assaf Yichia
    QA Manager
    Yokneam Illit
    Israel
    ------------------------------



  • 4.  RE: UDI for single-use implantable device

    Posted 20-Mar-2022 12:05
    And per section 12.1 (Implantable devices), "Implantable devices are generally not required to have a UDI carrier on the device itself (direct marking)".

    In the US, it is not required.

    For reference:
    https://www.federalregister.gov/documents/2013/09/24/2013-23059/unique-device-identification-system

    "Direct marking would generally serve no purpose as long as the device remains implanted, as there would be no way to read the direct marking except in those instances where RFID technology could be built into the device."

    ------------------------------
    Anne LeBlanc
    United States
    ------------------------------



  • 5.  RE: UDI for single-use implantable device

    Posted 21-Mar-2022 02:00
    Dear Anne,

    Thank you for your answer. It is very helpful. 
    Appreciate that.

    Best Regards,
    Assaf Yichia

    ------------------------------
    Assaf Yichia
    QA Manager
    Yokneam Illit