    Posted 19-Jul-2018 15:42
    During a recent audit, our AO indicated that we, as the manufacturer, must require all our suppliers to retain all their records for what they supply us either in accordance with jurisdictional requirements or transfer them to us after their internal record retention requirements are complete as required by ISO 13485:2016 clause 7.4.1 so that we can retain them for the life of the product. 

     

    However, when I read ISO 13485:2016 clause 7.4.1, to me it says that I must retain the supplier evaluation, selection, monitoring, and re-evaluation of the supplier in accordance with clause 4.2.5 but I do not see in the standard where it says that I must ask my suppliers to retain all their documents for as long as my medical device is on the market.

    I am wondering if anyone else has been asked to ensure that their suppliers conform to this requirement
    I would appreciate some input on this from the greater community



    ------------------------------
    Margaret Mucha (FRAPS) [Senior Leader RAQA]
    ------------------------------


  • 2.  RE: MDSAP/ISO 13485:2016 Supplier Document Retention question

    Posted 20-Jul-2018 05:24
    Hello Margaret,

    I have been asked the same or a couple similar type of instances by Notified Bodies/Registrars regarding supplier documentation.  This was especially true for a company I was helping that outsourced everything including having a Contract Manufacturer make the finished device.  While interpretation of Section 7.4.1 is specific the manufacturer must maintain those supplier records.  However, if you go back to Section 4.2.5 for quality records in general this would apply to anything related to QMS processes and specific to product records through product realisation, manufacturing, distribution, servicing, and post market.

    If you think about a relationship between the legal manufacturer of a medical device and the supplier there would need to be information in the contract, purchase order, or agreements that specify record retention requirements for those records the supplier generates.  In a less extreme example, you have a supplier that makes nuts and bolts for fastening parts in the assembly.  There probably would not be an expectation that supplier records would need to be retained for a long period of time or are least for a short time.  If a supplier manufactures a PCB board for a multiple use electronic device, then the records from the supplier would need to be retained for 'x' period of time.  What happens when there is servicing done, a customer complaint, or an issue with a lot of boards?  If the supplier destroys all the records after 6 months, how could a proper investigation be performed?  Then on most extreme example is a contract manufacturer of a finished medical device - this is still a supplier to the manufacturer.  It would be expected they would have the same record retention requirements as the legal manufacturer.  What happens during a recall, customer complaint, adverse event if the supplier does not have their records after one year?  Some suppliers in essence are extensions of the manufacturer which needs to follow the same requirements including record retention.

    While it may seem burdensome what the Notified Body is requesting, I have been asked this and do understand the rationale.  In fact, a few years ago we changed our standard supplier agreement to include specific requirements for record retention ... and not just the DHR.  The record retention was also including documents such as NCRs, Corrective Action, complaints, rework, etc.  It is not an expectation that the supplier must conform with ISO 13485 requirements (unless they are certified) but it is a requirement that the legal medical device manufacturer has the available records pre- and post-market just as they would need to keep the records.  Therefore this requirement gets passed along to the supplier.  Hope that helps gives some additional information.

    ------------------------------
    Richard Vincins RAC
    Vice President Regulatory Affairs
    ------------------------------



  • 3.  RE: MDSAP/ISO 13485:2016 Supplier Document Retention question

    Posted 20-Jul-2018 09:55

    With the possible exception of one case, I'm unable to find the requirement "indicated" by your AO. It is not in ISO 13485:2016, the companion document, or the MDSAP audit model.

     

    I infer that the AO issued a nonconformity. As such, it should have been written in the format prescribed by IMDRF/MDSAP WG/N24 FINAL: 2015 Medical Device Regulatory Audit Reports. Section 4.2.5 requires four elements:

    - identify the requirement against which the nonconformity is raised

    - make a statement of how that requirement has not been fulfilled

    - reference the supporting objective evidence in the audit summaries

    - assign the grade according to IMDRF/MDSAP WG/N3 – 9.1.2

     It would be valuable if you were to provide the wording of, at least, the first two requirements.

    You said, "require all our suppliers to retain all their records for what they supply us". This is, in my opinion an overreach when applied to "all" suppliers. This is the requirement I cannot find.

    There is a case when the supplier performs an outsourced process as described in 4.1.5. In that case it is your process performed at an external supplier's site so the record retention requirements for that process are your obligation.



    ------------------------------
    Dan O'Leary
    Swanzey NH
    United States
    ------------------------------



  • 4.  RE: MDSAP/ISO 13485:2016 Supplier Document Retention question

    Posted 23-Jul-2018 17:36
    Hi Dan
    This was the info on my NC
    Written against ISO 13485:2016 clause 7.4.1.  It was generically written.  I was told by our AO that the expectation is that we assume responsibility for the records from the supplier if they cannot maintain them for the lifetime of our device.

    Deviation:

    Supplier Quality Agreement are not always in compliance with the document retention and supplier evaluation/ re-evaluation criteria.

     Evidence: 

    Part 1:

    Purchase Agreement: Quality system documentation at the supplier's site must be retained for at least 5 years after the lifecycle of the products design or following manufacturing activities termination; manufacturing records shall be retained by the seller for at least 5 years after the lot/batch lifecycle.

    SOP Record Retention Requirement: Based on a lifetime of the medical device of 10 years.



    ------------------------------
    Margaret Mucha (FRAPS) [Senior Leader RAQA
    ------------------------------



  • 5.  RE: MDSAP/ISO 13485:2016 Supplier Document Retention question

    Posted 24-Jul-2018 07:59

    Thank you for providing this information. In my opinion, this goes into the examples of poorly written nonconformances.

    I agree that clause 7.4.1 is the correct clause for supplier management issues. However, I could make the case that this is also record retention.

    There are two ways to read the non-conformance.

    In the first way, the AO says that your purchase agreement does require record retention for 5 years and that you are not following your own procedure.

    In the second way, the AO says that your purchase agreement must have a provision for record retention for "at least 5 years after the lifecycle of the products design or following manufacturing activities termination". The assertion is that your procedure is deficient because it doesn't include this requirement.

    In my view, the AO had the second way in mind, which raises some issues.

    The first issue is the record retention period. ISO 13485:2016 clause 4.2.5 says, "The organization shall retain the records for at least the lifetime of the medical device as defined by the organization, or as specified by applicable regulatory requirements, but not less than two years from the medical device release by the organization."

    Since the nonconformance states the device life-cycle is 10 years, I don't understand where 5 years comes from, in terms of ISO 13485:2016.

    I looked at the MDSAP Companion Document. The Management Process, Task #8 covers record retention. It cites both Australian and Japanese record keeping requirements that include requirements with "5 years".

    The second issue is application to the QMS. I cannot find an ISO 13485:2016 requirement that says, in general, that a device manufacturer must require a supplier to maintain records based on the manufacturer's record retention. However, if there is an outsourced process, then purchasing and record retention would apply to that process.

    My conclusion is that is bad auditing. One of two things happened. Either the AO made up a "requirement" not supported by ISO 13485:2016 or the AO applied a country specific requirement without citing the requirement.

    ISO 13485:2016 has a potential ambiguity in that the record retention period is either the manufacturer's lifetime or the regulatory requirement with a floor of two years from device release. It doesn't say, "which ever is longer".

    My recommendation is that you ask the AO for the source of the "5 year" requirement. Ask them to withdraw the nonconformance if they it up or to rewrite the nonconformance to cite a country specific requirement. If so, you need to understand the country specific requirements, since it could include record retention by the supplier.



    ------------------------------
    Dan O'Leary
    Swanzey NH
    United States
    ------------------------------



  • 6.  RE: MDSAP/ISO 13485:2016 Supplier Document Retention question

    Posted 24-Jul-2018 08:34
    Thanks Dan
    I have asked now for the 3rd time the origin of the requirement specifically asking for the guidance document or other reference that gives this requirement.  Each time I get an explanation but not pointing to the actual requirement.  I will appeal this NC if I do not get further clarity.  This is the reason I posted this here-to see if others have had this type of NC and how they resolved it in case I was missing something.


    ------------------------------
    Margaret Mucha (FRAPS) [Senior Leader RAQA
    ------------------------------



  • 7.  RE: MDSAP/ISO 13485:2016 Supplier Document Retention question

    Posted 25-Jul-2018 04:25


    Hi Margaret:

    Hope you are doing well. 

    The requirements are embedded in the MDD Annex II clause 6 which require that records must  be maintained for inspection by the Competent Authorities for the "useful life of the product" or five years (10 for implants) from date of manufacture, whichever is greater.

    If your supplier generate records that are needed for product release e.g. sterilization you or your supplier need to make those records available to the CA.  

    I hope that helps.


    ------------------------------
    Armin Beck
    CEO
    SunTrixConsulting LLC
    El Dorado Hills, CA
    1-925-212-7683
    armingbeck@aol.com
    ------------------------------



  • 8.  RE: MDSAP/ISO 13485:2016 Supplier Document Retention question

    Posted 25-Jul-2018 04:41
    There seems to be a confusion in this discussion between suppliers (which supply product/service) and third parties (which acts on behalf of the manufacturer, as if they are the manufacturer, including complying with regulatory requirements in the name of the manufacturer).

    ------------------------------
    Marcelo Antunes
    Regulatory Strategy Consultant
    São Paulo
    Brazil
    ------------------------------



  • 9.  RE: MDSAP/ISO 13485:2016 Supplier Document Retention question

    Posted 25-Jul-2018 07:01
    The MDD is not in scope for an MDSAP audit. The EU does not participate (but they are an observer)..

    ------------------------------
    Dan O'Leary
    Swanzey NH
    United States
    ------------------------------



  • 10.  RE: MDSAP/ISO 13485:2016 Supplier Document Retention question

    Posted 25-Jul-2018 09:57
    Margaret,

    I believe what your auditor is stating is that some of your Supplier Agreements are not in line with your SOP, by view of their NC.  7.4.1 Purchasing processes speaks to evaluation and re-evaluation of your suppliers, 7.4.2 does point you in the right direction "To the extent required for traceability given in 7.5.9, the organization shall maintain relevant purchasing
    information in the form of documents (see 4.2.4) and records (see 4.2.5)." 

    Per 4.2.5 f and g.
    "f) ensure that documents of external origin, determined by the organization to be necessary for the
    planning and operation of the quality management system, are identified and their distribution
    controlled;
    g) prevent deterioration or loss of documents

    Which then filters to 4.2.5 Control of Records

    The organization shall retain the records for at least the lifetime of the medical device as defined by the
    organization, or as specified by applicable regulatory requirements, but not less than two years from
    the medical device release by the organization

    Your auditor may have been off a little on retention time. However his reference to 13485 seems correct.

    ------------------------------