Remember that inclusion of the safety and performance information on the website (if the manufacturer has a website) isn't optional; it is generally mandatory in all cases as appropriate (see my aforementioned decision tree). Also note that if the manufacturer doesn't also include the safety and performance information in the IFU, then the IFU will likely not conform with GSPR 23.4, as safety and performance information are mandatory elements of the IFU [for example, see 23.4(d) (where applicable), (e), (g), and others]. Accordingly, I would shy away from the notion of including safety and performance information on the website that isn't echoed in the IFU.
When deciding which particular safety and performance information is relevant to the user, or any other person, be aware of 23.1's language "as appropriate". This means that each manufacturer needs to, based on corresponding line item(s) in the risk analysis, decide which safety and performance information (if any) is relevant to the user, or any other person.
Hope this helps,
------------------------------
Kevin Randall, ASQ CQA, RAC (Europe, U.S., Canada)
Principal Consultant
ComplianceAcuity, Inc.
Golden, CO
United States
www.complianceacuity.comNote that I'm now far older and even uglier than in the photo above. Brace yourselves for my updated photo coming soon.
© Copyright 2020 by ComplianceAcuity, Inc. All rights reserved.
------------------------------
Original Message:
Sent: 12-Aug-2020 18:39
From: Maham Ansari
Subject: EU MDR on e-Labeling
Thank you all for such a great discussion on this. It has been immensely helpful. For my benefit, and also everybody else's on this Forum, I do have a follow-up question. If a manufacturer opts to include this safety and performance on the website rather than the IFU in order to comply with this GSPR, what would that information constitute? Are they expected to include all the residual risks, for instance? Any examples of what this could include would be great.
Thank you
------------------------------
Maham Ansari MS, RAC
Canada
Original Message:
Sent: 11-Aug-2020 16:19
From: Dan O'Leary
Subject: EU MDR on e-Labeling
Based on the information you provided, the audit nonconformance is correct.
The Annex I requirement is to provide certain information on the manufacturer's web site.
In your QMS you declare that the information is available in the IFU only.
Therefore, you must post the IFU on your website.
However, since this is an electronic IFU it must satisfy the 207/2012 requirements.
The solution (to quote a colleague of mine) is obvious to the most casual observer. Change your QMS requirement to allow the information on your web site but not in the IFU.
------------------------------
Dan O'Leary CQA, CQE
Swanzey NH
United States
Original Message:
Sent: 11-Aug-2020 05:16
From: Tze Yan Chia
Subject: EU MDR on e-Labeling
Hi Peter,
Thanks for pointing out that the 'information needed to identify the device and its manufacturer, and any safety and performance information relevant to the user' available on the website can be presented in many ways other than with the IFU.
In my situation, we declared that the information is only available in the form of IFU. That was the reason why the IFU has to be uploaded to the website and the requirement to comply with Regulation (EU) No 207/2012 is applicable.
------------------------------
Yan Chia
QARA Manager
Australia
Original Message:
Sent: 11-Aug-2020 04:35
From: Peter Reijntjes
Subject: EU MDR on e-Labeling
Hi Maham,
it is totally allowed to make the 'information needed to identify the device and its manufacturer, and any safety and performance information relevant to the user' available on the website in other ways than with the IFU. And in that situation the requirement to comply with Regulation (EU) No 207/2012 on electronic labeling is not applicable. But this is rather theoretical and many manufacturers will decide to use the IFU as device identification safety and performance information to be uploaded.
I am however a bit worried about the statement from Yan that the Notified Body required the eIFU has to be uploaded to the website, taking into account above mentioned reasoning. But not knowing the situation and arrangements made in your QMS it is difficult to comment on this any further.
------------------------------
Peter Reijntjes
Principal Consultant Regulatory & Quality Affairs | Head of Training
Arnhem
Netherlands
Original Message:
Sent: 11-Aug-2020 02:07
From: Tze Yan Chia
Subject: EU MDR on e-Labeling
Hi Maham,
That is correct. I have just completed the MDR audit and one of the comments from the notified body was the eIFU has to be uploaded to the website (even when paper copies are provided) and it has to be Regulation (EU) 207/2012 compliant.
As a result, the risk assessment document and SOP to prepare IFU demonstrating EU 207/2012 compliant are required, these documents should be referenced in the GSPR checklist.
Hope it helps.
------------------------------
Yan Chia
QARA Manager
Australia
Original Message:
Sent: 10-Aug-2020 17:00
From: Maham Ansari
Subject: EU MDR on e-Labeling
Within the new EU Medical Device Regulations (EU) 2017/745, General Safety and Performance Requirement (GSPR) 23.1 states the following:
"Each device shall be accompanied by the information needed to identify the device and its manufacturer, and by any safety and performance information relevant to the user, or any other person, as appropriate. Such information may appear on the device itself, on the packaging or in the instructions for use, and shall, if the manufacturer has a website, be made available and kept up to date on the website.." [emphasis added]
Should this requirement be interpreted to mean that the IFU must be available on the manufacturer's website (if they have one), or is this still optional? If so, it would also now become obligatory to comply with Regulation (EU) No 207/2012 on electronic labeling (Article 9 in particular)?
Does the MDR essentially mandate supply of electronic IFUs, even when paper copies are provided, just because a manufacturer has a website, which majority of the manufacturers do?
Thank you.
------------------------------
Maham Ansari MS RAC
Director, Regulatory Affairs
Canada
------------------------------